- | ===== Fluorescence in UV/Vis ===== | ||
- | ==== Standard Emission and Excitation Spectra ==== | ||
- | Instructions that must be followed to maintain the integrity of the fluorimeter components are indicated in red. | ||
- | |||
- | === INITIATE INSTRUMENT === | ||
- | |||
- | **1)** Acquaint yourself with the instrument layout and the light path to the detectors using the labeled pictures. | ||
- | |||
- | **2)** Turn on components in following order. | ||
- | * a. Main fluorimeter. | ||
- | * b. Lamp. Located directly above Main switch. | ||
- | * c. Dell computer. | ||
- | * d. iHR 320 monochromator. | ||
- | * e. SpectraAco computer. | ||
- | * f. Symphony power supply. Press button on lower right hand corner of its front face. | ||
- | **3)** Allow at least 20 minutes for the lamp to warm up. (For best results; measurements can be taken prior but the lamp spectrum may fluctuate) | ||
- | |||
- | **4)** The Operation Manual and the Test Spectra, which will be needed for calibration, | ||
- | |||
- | === CHECK CALIBRATION OF MONOCHROMATORS === | ||
- | |||
- | **5)** At computer, open the FluorEssenceV3 program on the desktop. | ||
- | |||
- | **6)** Click on M icon (Experiment Menu) (Figure 1). The various devices will be checked for their connection to the computer. | ||
- | |||
- | **Figure 1** | ||
- | |||
- | {{: | ||
- | |||
- | === EXCITATION === | ||
- | (See pages 3-4 to 3-9 of Operation Manual.) | ||
- | |||
- | **7)** From the Main Experiment Menu, choose Spectra (Figure 2). | ||
- | |||
- | **Figure 2** | ||
- | |||
- | {{: | ||
- | |||
- | **8)** Select Experiment Type: Excitation. | ||
- | |||
- | **9)** The Experiment Setup window opens with the “Monos” menu (see left) open (Figure 3). The IHR320 is not needed for these measurements, | ||
- | |||
- | **Figure 3** | ||
- | |||
- | {{: | ||
- | |||
- | **10)** Click on the “Detectors” menu on left. The Experimental Setup window should look like Figure 4. The R1 detector (which is a reference detector that does not have any monochromators in front of it) and the S1 detector (which is the PMT on the far side of the table and is for these basic measurements) should be Enabled. Verify the shutter on the PMT detector is open (fully extended) before running scan. | ||
- | |||
- | **Figure 4** | ||
- | |||
- | {{: | ||
- | |||
- | **11)** Click Run. The data is collected and should look like that in Figure 5. The data from the R1 detector should look like the lamp spectrum in the “Test Spectra” booklet. | ||
- | |||
- | **12)** Click on the Data Reader icon and click on the plot from the R1 detector. Move the crosshairs to the maximum. | ||
- | |||
- | **Figure 5** | ||
- | |||
- | {{: | ||
- | |||
- | === EMISSION === | ||
- | (See pages 3-10 to 3-12 of Operation Manual.) | ||
- | |||
- | **13)** You use the water Raman spectrum to calibrate the emission monochromator. | ||
- | |||
- | **14)** The calibration occurs in the Emission 2 tab of the Monos section of the Real Time Control panel. | ||
- | |||
- | **Figure 6** | ||
- | |||
- | {{: | ||
- | |||
- | === MAKE MEASUREMENTS === | ||
- | |||
- | **15)** With the excitation and emission monochromators calibrated, you can now take measurements. | ||
- | |||
- | **16)** There are other adjustments that can be made to optimize data collection (see Chapter 5: Optimizing Data of Operation Manual). | ||
- | * a. Using a grating with a blaze wavelength (the second number in the choices in the grating menu) suitable for your sample | ||
- | * b. Integration time (shorter for quick measurements to see what’s going on/bright samples and longer (0.5 s and longer) for publication quality measurements/ | ||
- | * c. Saving the Experiment file to make future measurements with same settings easier | ||
- | * d. Measurements can be performed in sequence using the Batch Experiment Button (three buttons to the right of the Experiment Menu button). | ||
- | |||
- | === DATA ANALYSIS AND SHUTTING DOWN === | ||
- | |||
- | **17)** Once you’re done, you can save the day’s measurements as a project to open up again in Origin. | ||
- | * **!!Shut down the FluorEssenceV3 Software before shutting off any hardware!! Turn off all machines in the reverse order of startup EXCEPT FOR THE MAIN FLUORIMETER. | ||
- | |||
- | **18)** Clean off the top of the desk. Put these instructions and the Operation Manual and Test Spectra in the top left drawer beneath the printer. |